Emerson Process Management ROSEMOUNT 5300 Manual do Utilizador Página 116

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 324
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 115
Reference Manual
00809-0100-4530, Rev BA
July 2009
Rosemount 5300 Series
5-44
Application Example 3 Radar Level Transmitter, Product Level and Interface Level Value
A level transmitter is measuring the product level and the interface level in a
33 ft. (10 m) high tank. The maximum interface level is 10 ft. (3 m).
Figure 5-16. Situation Diagram
Solution
Table 5-11 lists the appropriate configuration settings, and Figure 5-17
illustrates the correct function block configuration.
Table 5-11. Analog Input
Function Block Configuration for
a Level and Interface
Transmitter
Figure 5-17. Analog Input
Function Block Diagram for a
Level and Interface Transmitter
10 ft
(3 m)
100%
0%
33 ft
(10 m)
100%
0%
AI Function Block Product Level AI Function Block Interface Level
Parameter Configured Values Parameter Configured Values
L_TYPE Direct L_TYPE Direct
XD_SCALE Not Used XD_SCALE Not Used
OUT_SCALE Not Used OUT_SCALE Not Used
CHANNEL CH1: Level CHANNEL CH11: Interface Level
Level Measurement
To Another
Function Block
OUT_D
OUT
AI Function Block
AI Function Block
(Product Level)
To Another
Function Block
To Another
Function Block
OUT_D
OUT
AI Function Block
AI Function Block
(Interface Level)
To Another
Function Block
Interface Level Measurement
Vista de página 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 323 324

Comentários a estes Manuais

Sem comentários