Reference Manual
00809-0100-4530, Rev BA
July 2009
C-5
Rosemount 5300 Series
NOTE!
Note that for firmware version 2.A2 the Trim Near Zone is not possible in
combination with Narrow Nozzles, as defined below:
50 mm< Nozzle height <
300 mm
Nozzle diameter <
2 in. for all single probes except 13 mm single rigid
Nozzle diameter < 3 in. for 13 mm single rigid
Figure C-3. Trim Near Zone
using RRM.
1. Click Trim Near Zone.
2. Select Trim near zone and click OK.
3. Click OK in the appearing dialog.
4. Wait 1 minute.
5. Restart the transmitter.