Reference Manual
00809-0100-4530, Rev BA
July 2009
C-21
Rosemount 5300 Series
Figure C-18. Check if Signal
Quality Metrics is supported.
3. If “Diagnostics Suite” is mentioned in the Device Software
Configuration 2 list, the device supports Signal Quality Metrics.
In 375 Field Communicator, if Signal Quality Metrics is supported, it can
be found with the [3, 2, 2, 1] sequence. Check if “Diagnostics Suite” is
present
Signal Quality Metrics can be enabled/disabled in RRM. Choose Tools,
Advanced, Signal Quality Metrics tab.
NOTE!
If Signal Quality Metrics is not supported or disabled, the Signal Quality and
Surface / Noise Margin will always be set to 0.
Viewing Signal Quality
Metrics in RRM
To view Signal Quality Metrics in RRM, choose the Tools > Device Display
option and select the Signal Quality Metrics tab.
Figure C-19. View Signal Quality
Metrics Values.